XS user manual v0.55 - 102 -
the « L » letter is followed by a number between 1 and 8 , giving the test step number and
then by a « space ».
then comes a letter according to the type of the test step :
Function
French
English
German
Insulation
M
M
M
Dielectric strength
R
H
H
Continuity
C
G
E
Void
. .
. .
. .
Pause
OK.
OK.
OK
the measurement type is followed by the memory number which contains the
measurement parameters , then by a separator character « : »
if the test step is void , the measurement will be replaced by a « space »
otherwise , the next character gives the test step result :
« space » indicates that the test is good
« q » indicates that the test is bad
for the insulation and dielectric strength tests the previous character is followed by 1
‘space’ and for the ground continuity test by 2 ‘space’.
then the numerical values are given . When the value cannot be measured ( no continuity
or overload ) the numerical indication is replaced by « ---- »
To check if a tested specimen is good or bad at the end of a sequence , it is necessary to
check the STB register ( bit 3) .To get the result of each test step , check if the result
character is « space » or « q ».