XS series operating manual 
 
XS user manual v0.55  - 102 - 
 
 
  the « L » letter is followed by a number between 1 and 8 , giving the test step number and 
then by a « space ». 
  then comes a letter according to the type of the test step : 
 
  the measurement type is followed by the memory number which contains the 
measurement parameters , then by a separator character « : » 
  if the test step is void , the measurement will be replaced by a « space » 
  otherwise , the next character gives the test step result : 
  « space » indicates that the test is good  
  « q » indicates that the test is bad 
  for the insulation and dielectric strength tests the previous character is followed by 1 
‘space’ and for the ground continuity test by 2 ‘space’. 
  then the numerical values are given . When the value cannot be measured ( no continuity 
or overload ) the numerical indication is replaced by « ---- » 
 
To check if a tested specimen is good or bad at the end of a sequence , it is necessary to 
check the STB register ( bit 3) .To get the result of each test step , check if the result 
character is « space » or « q ».