XS series operating manual
XS user manual v0.55 - 148 -
Uses cases
Test piece (DUT) correctly connected.
Reminder:
: Voltage re-reading lead.
: High voltage generator HV lead.
: Voltage re-reading return lead.
: Return lead (GND) from the high voltage generator.
When the test piece is properly connected, points 1 and 2 are connected together via connection zone A.
The same applies to points 3 and 4 connected together via connection zone B. In this case the high
voltage is returned to the voltage shunt and the other terminal is brought to 0V (GND). Relay Re1
contacts are open.
Hence the voltage generated can be re-read across the voltage shunt.
DUT
1
2
3
4
Conduction
zone A
Conduction
zone B
2
3
4
1