Page 34
ecom-B
-calculated with reference O
2
– formula for the calculation:
O
2
Reference (set by pressing <Enter>):
-setting of reference oxygen value O
2ref
-reset by pressing <CAL>
Air pressure (set by pressing <Enter>):
-Input of barometric pressure for calculation of dew point
-Reset by pressing <CAL>
Keyboard beep (set with cursers <up / down>)
-set signal when hitting the keys
-reset by pressing <CAL>
AUX(COM) (adjustment via <CAL.>):
-Adjustment of transfer speed by data transfer via RS 232
(1200 -112500 Baud)
Online data (adjustment via cursor keys):
-Adjustment what kind of Data should be sent:
-No protocol: Data for program „miniDV“
-Remote: only for special applications
-DAS: Data for program „DASNT“ (1200 Baud)
The programs „miniDV“ und „DASNT are available free of charge
from the ecom website.
Bluetooth (option)
(<BLE> no adjustment needed):
-Online data = Remote
E
ref
= E
meas
*
21 – O
2ref
21 – O
2meas