This manual is the intellectual property of Ecoso. Copying and reprinng is prohibited. © 2021
21
INSTRUCTION MANUAL ECOSOFT RO SYSTEM
MO6500BWTSS
2. Calibraon
Hold STOP for 8 seconds to launch menu prompt. Push STOP to skip Sengs submenu and
push
START to enter Calibraon submenu. Enter passcode in the prompt. Factory passcode is
‘0000’.
2.1 First point value
First calibraon point can be done at zero electrical conducvity (dry conducvity meter). In order
to use zero rst point conducvity, remove the conducvity meter from its cell, wipe with clean
cloth and keep dry for a few minutes. When conducvity reading on the display stabilizes, put
zeroes in the boom row, and go to the next step.
If using a weakly conducng soluon to set the rst point, rinse the meter with deionized water
and wipe dry. Dip clean conducvity meter into sample of known standard conducvity, wait unl
the reading on display stabilizes and input actual conducvity. Then go to the next step.
2.2 Second point value
Use water sample with greater conducvity than that of the rst point standard. Follow the
same procedure rinsing and wiping residual moisture on conducvity meter electrodes. Dip clean
conducvity meter into sample of known standard conducvity, wait unl the reading on display
stabilizes and input actual conducvity. Then go to the next step. The controller will display ‘OK’
and show Maintenance submenu prompt.
3. Maintenance
Maintenance submenu will be shown aer compleng calibraon of conducvity meter and can
be called up during Service by holding STOP for 8 seconds, then skipping Sengs and Calibraon
prompt displays. Enter Maintenance passcode in the prompt. Factory passcode is ‘0000’.
3.1 Schedule maintenance stop
Select ‘on’ to turn on maintenance reminder aer preset number of hours of cumulave runme.
Controller will put the RO machine to a halt and display maintenance alert message. Operaon can
only be connued aer entering Maintenance submenu (with proper Maintenance passcode) and
reseng scheduled stop period. If set to ‘o’, the controller will connue to count overdue hours
aer reaching zero hour count.
3.2 Scheduled stop period
Enter the number of hours before the RO machine will be brought to a scheduled stop for
maintenance. This seng will not be shown if the scheduled stop is turned o in step 3.1.
3.3 New Maintenance passcode
Enter new passcode for Maintenance submenu and conrm. This will exit the Conguraon menu.