FTS-3 Fetal Telemetry System Service Manual Installation
- 8 -
Chapter 3 Installation
WARNING
Only qualified service engineers should install this equipment.
3.1 Environment Requirements
Environment
Working
+5 ºC ~ +40 ºC ( +41ºF ~ +104ºF)
15% ~ 93% (non-condensing)
86 kPa ~ 106 kPa
Transport
and Storage
-20 ºC ~ +55 ºC (-4ºF ~ +131 ºF)
15% ~ 93% (non-condensing)
70 kPa ~ 106 kPa
Classification
Anti-electric Shock Type:
Class I equipment with internal power supply
Anti-electric Shock Degree:
FHR1, FHR2, TOCO BF
DECG, MHR(form MECG) CF
Transducer Degree of Protection
against Harmful Ingress of Water:
IPX8 (Protected against the effects of continuous
immersion in water 1.1m deep for 24 hours)
Degree of Safety in Presence of
Flammable Gases:
Equipment not suitable for use in presence of
flammable gases
Disinfection/Sterilizing Method:
Refer to User Manual for details
EMC:
CISPR11 Group 1 Class A
Working System
Continuous running equipment
NOTE:
1. Ensure FTS-3 is not subject to any source of strong electromagnetic interference, such
as CT, radio transmitters, mobile phones base stations, etc.
2. Do not install FTS-3 in a flammable atmosphere where concentrations of flammable
理邦保密文件
EDAN CONFIDENTIAL