FTS-3 Fetal Telemetry System Service Manual Periodic Tests
- 40 -
Figure 11-3
4) Check that the value on the display shows this change in pressure.
If a TOCO-T or TOCO-E transducer fails the test, repeat this test with another transducer. If the
second one passes the test, defect of the first transducer is confirmed. Replace it with a good one.
If the second transducer fails the test as well, check the monitor and FTS-3 system.
11.2.3 ECG Performance Test
This test checks the MHR performance of the TOCO-E transducer.
Required equipment: FLUKE PS320 simulator.
1) Connect the FTS-3 base station to the monitor and switch on them.
2) Take the TOCO-E transducer from the base station and the transducer will switvh on.
Use the ECG cable to connect the transducer and the simulator as shown below.
Figure 11-4