FTS-3 Fetal Telemetry System Service Manual Disassembling FTS-3
- 33 -
10.4 Disassembling the Upper assembly
10.4.1 Replacing the Charging Interface Board
To remove the charging interface board,
(1) Separate the upper assembly and the lower assembly.
(2) Turn on the connector swich with both index fingertips and remove the adhesive tape.
Figure 10- 9
(3) Disconnect the wire.
(4) Remove four screws securing the charging interface board.
To install the charging interface board,
Install the charging interface board in a reverse procedure. Refer to the following illustration
when connecting the wires. Ensure it is correctly connected.
Figure 10- 10