Vital Signs Monitor User Manual Product Specification
- 139 -
Nellcor Module
Measuring Range 1% to 100%
Resolution 1%
Data Update Period
1 s
Accuracy
DS-100A, OXI-A/N (Adult)
D-YS (Adult and Pediatric)
OXI-P/I (Pediatric)
± 3% (70% to 100% SpO
2
)
MAX-A, MAX-AL, MAX-N, MAX-P,
MAX-I, MAX-FAST (Adult and
Pediatric)
±2% (70% to 100% SpO
2
)
MAX-A, MAX-AL, MAX-N, MAX-P,
MAX-I, MAX-FAST (Adult and
Pediatric)
±3% (60% to 80% SpO
2
)
If sensor is used for neonate as recommended, the accuracy will be larger than
adult by ±1.
SpO
2
Storage
Interval
In Ward round or spot-checking mode
30 S (default), 1 min, 2 mins, 5 mins
Sensor Wave length: approximately 660 nm and 900 nm
Emitted light energy: <15 mW
NOTE:
The information about wavelength range can be especially useful to clinicians (for
instance, when photodynamic therapy is performed).
A.5 PR
Measuring range Accuracy Resolution
PR (SpO
2
) EDAN 25 bpm to 300 bpm
±
2 bpm
1 bpm
Nellcor 20 bpm to 300 bpm
±
3 bpm (20 bpm to 250 bpm)
1 bpm
PR (NIBP) EDAN 40 bpm to 240 bpm ±3 bpm or 3.5%, whichever is
greater
1 bpm
SunTech 30 bpm to 220 bpm ±3 bpm or ±2%, whichever is
greater
1 bpm