Vital Signs Monitor User Manual Alarms
- 52 -
Remind signal: Audio alarm off symbol and Audio Alarm off on a red colored background
are displayed with an interval of 2 s during the audio alarm off status.
Pressing shortcut key again can resume the audio alarm.
NOTE:
If a new alarm occurs during the audio alarm off period, the new alarm will not be
sounding.
8.3.4 Alarm Reset
In monitor mode, select the shortcut key on the screen directly. When the alarm is
reset,
No alarms are sounding until a new alarm occurs.
As for the active alarms, the visual alarm indications are still displayed.
All latching alarms are cleared. If the alarm condition is no longer present, all alarm
indications stop and the alarm is reset.
It will not influence the configuration of physiological alarm off, audio paused, and audio off
status.
NOTE:
If a new alarm occurs after the alarm is reset, the new alarm will be sounding.
8.4 Latching Alarms
In monitor mode, to configure the alarm latching setting, select Menu > Maintenance > User
Maintain > Alarm and choose Alarm Latch which can be set to On or Off. When it is set to
Off, alarm indications end when the alarm condition ends. When it is set to On, the visual and
audio alarm indications are still displayed after the alarm condition ends; meanwhile, the alarm
time is also displayed for the latched alarm for your reference. The indication lasts until you
acknowledge the alarm.
You can use the permanent key on the screen to acknowledge the latched alarm.
8.5 Alarm of SpO
2
Sensor Off
In monitor mode, user can enter Menu > User Maintain > Alarm to set the alarm level of SpO
2
Sensor Off to High, Med. or Low. It is set to Low by default.
In Ward round or spot-checking mode, it is acted as prompt and the level is not user-selectable.