Vital Signs Monitor User Manual Product Specification
- 140 -
A.6 TEMP
Complies with ISO 80601-2-56: 2017+A1:2018.
T2A Module:
Measuring range
Monitor mode: 25 °C~45 °C
Predict mode: 35.5 °C~42 °C
Working temperature
10 °C ~40 °C
Transport and Storage
-20 °C ~ 55 °C
Sensor type Oral /axillary /rectal
Adjustable range of alarm limits
35.5 °C ~42 °C
Resolution
0.1 °C
Accuracy
Monitor mode: ±0.1 °C (25 °C~ 45 °C)
Response time < 60 s
Update time 1 s~ 2 s
Warm-up time <10 s
Time for predicting <30 s
Calibration Self-test interval: 5 min
Measuring Mode Direct Mode/ Adjusted Mode
Transient Response Time 30 s
Clinical Bias
(-0.2 to -0.4 )°C
Limits of Agreement 0.49
Stability
0.14 °C
NOTE:
The direct mode refers to monitor mode, while adjusted mode refers to predict mode.
TH Module:
Measuring range
34 °C~ 42.2 °C
Working temperature
10 °C~ 40 °C
Transport and Storage
-20 °C ~ 50 °C
Adjustable range of alarm limits
35.5 °C~ 42 °C
Resolution
0.1 °C
Response time 1 s