Vital Signs Monitor User Manual Product Specification
- 141 -
Measuring Mode Adjusted Mode
Clinical Accuracy
0.2 °C (0.4 °F) (35.5 °C ~42 °C) (95°F
~107.6°F)
±0.3 °C (0.5°F) (out of the range mentioned above)
Laboratory Accuracy
0.2 °C
F3000 Module:
☆Measuring range
30 °C~43 °C
Prediction measurement range
35 °C~43 °C
Low Temp. Mode prediction
measurement range
33 °C~43 °C
Working temperature
10 °C ~ 40 °C
Transport and Storage
-20 °C ~ 55 °C
Sensor type Oral /axillary /rectal
☆Adjustable range of alarm limits
35.5 °C ~42 °C
Resolution
0.1 °C
☆Accuracy
Monitoring Mode and Preditive Mode: ±0.1 °C
Quick Predictive Mode: ±0.3 °C
Typical measurement time (after
insertion into measurement site)
Oral (Quick Predictive Mode): (3~5) s (non-fever
temps); (8~10) s (fever temps)
Oral (Predictive Mode): (6~10) s
Axillary: (8~12) s
Rectal: (10~14) s
Monitoring Mode (all sites): (60~120) s
Measuring Mode Direct Mode /Adjusted Mode
Transient Response Time
≤30 s monitoring mode
Clinical Bias
(-0.2 to -0.4 )°C
Limits of Agreement 0.49
Stability
0.14 °C
NOTE:
The direct mode refers to monitor mode, while adjusted mode refers to predictive mode
and quick predictive mode.