Vital Signs Monitor User Manual Alarm Information
- 66 -
SpO
2
Desat Limits are listed as follows (unit %):
Adjustable Range
SpO
2
Desat Limit
20~99
NOTE:
User can set the range through User Maintain > Alarm > SpO
2
Desat Limit. SpO
2
Desat Limit should be less than or equal to SpO
2
alarm low limit.
PR
alarm limits are listed as follows: unit (bpm)
Adjustable Range
PR (SpO
2
) EDAN 30~300
Nellcor
30~300
PR (NIBP) EDAN
40~240
SunTech 30~220
NIBP alarm limits are listed as follows:
EDAN module:
Patient Type Adjustable Range (mmHg)
ADU SYS
DIA
MAP
40~270
10~215
20~235
PED SYS
DIA
MAP
40~230
10~180
20~195
NEO SYS
DIA
MAP
40~135
10~100
20~110
SunTech module:
Patient Type Adjustable Range (mmHg)
ADU SYS
DIA
MAP
40~260
20~200
26~220
PED SYS
DIA
MAP
40~230
20~160
26~183
NEO SYS
DIA
MAP
40~130
20~100
26~110