iSE Series Electrocardiograph Service Manual Inspection and Troubleshooting
- 30 -
5. Parameter Troubles
Problem Description
Possible Causes
Actions
No ECG waveform
Poor connection between
the patient and the
electrodes
Poor connection between
the ECG cable and the
electrocardiograph
Bad connection cable of
parameter board
ECG board or main
board failure
Attach the electrodes to the
patient again or clean the
electrode location on body
surface with alcohol
Connect the ECG cable to the
electrocardiograph
Replace the connection cable
Replace the ECG board or
main board
ECG waveform is
abnormal or disturbed
Poor connection between
the patient and the
electrodes
ECG cable failure
ECG board failure
Attach the electrodes to the
patient again or clean the
electrode location on body
surface with alcohol
Replace the ECG cable that is
defective
Replace the ECG board
Heart rate is not accurate
Waveform measuring failure
Adjust the electrode connection to
the body
ECG waveform has burrs
Poor connection between
the patient and the
electrodes
ECG board failure
Electrical interference
from another device
(microwave oven,
cellular phone, wireless
device, etc.)
An improperly-grounded
electrical device near the
Attach the electrodes to the
patient again or clean the
electrode location on body
surface with alcohol
Replace the ECG board
Look for devices that could be
causing electrical interference,
and then unplug the devices.
Or run the electrocardiograph
on the battery power.
Lay the lead wires alongside
the limbs and away from any