iSE Series Electrocardiograph Service Manual Inspection and Troubleshooting
- 29 -
4. Operation Troubles
The touch screen is damaged.
Replace the touch screen or upper
casing parts.
The horn sound is hoarse
or broken.
The electrocardiograph
cannot print reports.
① The recorder paper runs
out.
② Print head failure
③ Main board or boost
board or printer driver
failure
④ Recorder wiring fault
① Load the recorder paper
② Replace the printer module
③ Replace the main board or
boost board or printer driver
④ Replace or repair the recorder
wiring
The printed report is
illegible
Some characters or
waveforms printed are
missing.
① The rubber roller of the
recorder casing has stain.
② Print module failure
③ Dirty thermal printer head
④ The recorder frame is not
good.
① Clean the rubber roller
② Replace the print module
③ Clean the thermal printer head
④ Replace the recorder frame
① There is a stain on the
detecting position of the
printing bracket.
② Reflective photo sensor
failure
① Clean the stain on the
detecting position of the
printing bracket
② Replace the reflective photo
sensor