iSE Series Electrocardiograph Service Manual Inspection and Troubleshooting
- 29 -
4. Operation Troubles
Problem Description
Possible Causes
Actions
Touch screen failure
The touch screen is damaged.
Replace the touch screen or upper
casing parts.
The horn sound is hoarse
or broken.
Speaker failure
Replace the speaker
The electrocardiograph
cannot print reports.
The recorder paper runs
out.
Print head failure
Main board or boost
board or printer driver
failure
Recorder wiring fault
Load the recorder paper
Replace the printer module
Replace the main board or
boost board or printer driver
Replace or repair the recorder
wiring
The printed report is
illegible
Some characters or
waveforms printed are
missing.
The rubber roller of the
recorder casing has stain.
Print module failure
Dirty thermal printer head
The recorder frame is not
good.
Clean the rubber roller
Replace the print module
Clean the thermal printer head
Replace the recorder frame
Fail to detect paper
There is a stain on the
detecting position of the
printing bracket.
Reflective photo sensor
failure
Clean the stain on the
detecting position of the
printing bracket
Replace the reflective photo
sensor