iSE Series Electrocardiograph Service Manual Inspection and Troubleshooting
- 28 -
main board or the
electrocardiograph
2. Display Troubles
Problem Description
Possible Causes
Actions
Some characters displayed on
the LCD screen are missing.
LCD screen failure.
Replace the LCD screen
The LCD screen is dim.
The brightness of the
LCD screen is not
adjusted well.
LCD screen failure
Adjust the brightness of the
LCD screen in the
notification bar.
Replace the LCD screen
3. Transmission Troubles
Problem Description
Possible Causes
Actions
Fail to transmit ECG data
through the network
Fail to copy data from the
electrocardiograph to the
U disk or SD card
Setup error
LAN cable failure
USB interface or LAN
port failure
USB storage device
failure
SD card failure
Examine whether the
transmission settings are
correct.
Replace the LAN cable
Replace the USB interface or
LAN port
Replace with a new USB
storage device provided or
recommended by the
manufacturer
Replace with a new SD card
provided or recommended by
the manufacturer
SD card slot on the main
board is damaged
Fail to connect to the
network or to transmit
ECG data through
4G/WiFi/BT/NFC/GPS
Setup error
Antenna failure
Software installed in PC is
abnormal
Main board failure
Examine whether the
transmission settings are
correct.
Replace the antenna
Restart the software in your
PC
Replace the main board