3. Technical data
3.1 Oper
ang and storage condions
Table 3
Operang and storage condions
Range Data
Ambient operang temperature range 5 °C to 40 °C
Ambient operang humidity range 80% up to 31 °C
Reducing linearly to 50% at 40 °C
Maximum operang altude 3000 m
Ambient storage temperature range -30 °C to 70 °C
Maximum humidity (storage in original packag-
ing)
≤ 95 % RH
Equipment type Indoor use
Enclosure protecon (tested according to IEC
60529)
IP21
Enclosure pr
otecon (tested according to UL50e
and CSA C22.2 No. 94.2)
NEMA 1
Table 4
Environmental condions
Polluon Polluon degree 2
Installaon Installaon category II
Altude restricon Maximum 3000 m
Area of use Indoor
Table 5
Mechanical data
Parameter
nXR30i nXR40i nXR60i nXR90i nXR120i
Mass 27.6 kg 29.6 kg 30 kg
Inlet c
onnecon NW25 NW40
Overall dimensions (L x W x H) 478x217x301 494 x 217 x 301 mm
Maximum lt angle for operaon ± 5°
Maximum lt angle for transport 30°
Nominal rotaonal speed 15000 rpm
Outlet connecon NW25
05/2021 - ©Edwards Limited
Page 13A90301880_C
A90301880_C - Technical data