temperature or di
erent ambient pressures pump performance may be compromised
and the rotaonal speed may drop below nominal.
Note:
F
or the nXR120i, the pump may thermally protect itself and shut down, if:
1) the connuous operang pressure at the inlet is greater than 20 mbar
2) the nXR120i is used to evacuate large chambers, parcularly if large connuous gas
loads are combined with the chamber pump-down, Table: nXR120i applicaon guidelines
for chamber pump-down provides guidelines on the acceptable applicaon parameters.
It is also advised to start chamber evacuaon with the pump at rest than with the pump
already operang at full rotaonal speed.
Figure 5 nXR120i applicaon guidelines for chamber pump-down
0
2
4
6
8
10
12
14
16
18
20
0 50 100 150 200 250 300 350
Gas flow (slm)
Chamber volume (litres)
GE/3806/A
Table 8
Sound data
Declared dual-number noise emission values in accordance with ISO 4871
nXR30i nXR40i nXR60i nXR90i nXR120i
Measured A-weighted emission sound pres-
sure level, L
pA
at ulmate vacuum 1 m from
the pump in free space
55.0 dB(A)
Measured A-weighted emission sound pres-
sur
e level, L
pA
at 15 mbar at inlet 1 m from the
pump in free space
57.0 dB(A)
Uncertainty, K
pA
2.5 dB(A)
Product tested to ISO standard 3744:2010
05/2021 - ©Edwards Limited
Page 15A90301880_C
A90301880_C - Technical data