6 Operation
Page 26 / 42
check conformance of the HyperLight-G. Using other test instruments may
yield differing results.
Finally, IBA sensors will be used for performance testing. Compare the
results with the factory release parameters (indicated in the chart below). For
results outside these parameters, discontinue use and contact your
dealer/distributor or Sifary.
Test
Descriptio
n
Accepta
nce
Limits
Timer Settings and Corresponding Acceptable
Ranges
0.05s
0.1s
0.5s
1s
2s
kVp (eff)
Accuracy
65kV±10
%
58.5-71.
5 kV
58.5-71.
5 kV
58.5-71.
5 kV
58.5-71.
5 kV
58.5-71.
5 kV
Timer
Accuracy
Setpoint
±(5%
or
20ms
0.044-0.
056s
0.091-0.
111s
0.451-0.
551s
0.901-1.
101s
1.801-2.
201s
Tube
Current
2 mA
(±20%)
1.6-2.4
mA
1.6-2.4
mA
1.6-2.4
mA
1.6-2.4
mA
1.6-2.4
mA
Dose rate
1.6-3
mGy/s
1.6-3
mGy/s
1.6-3
mGy/s
1.6-3
mGy/s
1.6-3
mGy/s
1.6-3
mGy/s