11 EMC Tables
Page 36 / 42
11 EMC Tables
Guidance and manufacturer’s declaration electromagnetic immunity
The HyperLight-G is intended for use in the electromagnetic environment specified below.
The customer or the user of the HyperLight-G should assure that it is used in such an
environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic
environment - guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
+/-8 kV contact
+/-15kV air
+/-2, 4, 6 & 8kV
contact
+/-2, 4, 8, & 15 kV
air
Floors should be wood,
concrete or ceramic tile. If
floors are synthetic, the r/h
should be at least 30%
Electrical fast
Transient/burst
IEC 61000-4-4
+/-2 kV for
power supply
lines
+/-1 kV for input/
output lines
+/-0.5, 1 & 2 kV
for power
supply lines
+/-0.5 & 1 kV for
input/output lines
Mains power quality should
be that of a typical
commercial or hospital
environment.
Guidance and manufacturer’s declaration electromagnetic emissions
The
HyperLight-G
is intended for use in the electromagnetic environment specified below.
The customer or the user of the HyperLight-G should assure that it is used in such an
environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions CISPR 11
Group 1
The HyperLight-G uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emissions CISPR 11
Class B
The
HyperLight-G
is suitable for use in all
establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC61000-3-2
Class A
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Complies