11 EMC Tables
Page 37 / 42
Surge
IEC 61000-4-5
+/-1 kV
differential mode
+/-2 kV common
mode
+/-0.5 & 1 kV
differential mode
+/-0.5, 1 & 2 kV
Mains power quality should
be that of a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
100 % Dip; for
0,5 cycles
At 0°, 45°,
90°, 135°,
180°, 225°,
270° and 315°
60% Dip for
1 Cycle
30% Dip for
25 Cycles
100% Dip for
5 Seconds
100 % Dip for 0,5
cycles
At 0°, 45°,
90°, 135°,
180°, 225°,
270° and 315°
60% Dip for
1 Cycle
30% Dip for
25 Cycles
100% Dip for
5 Seconds
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the
HyperLight-G
requires continued operation
during power mains
interruptions, it is
recommended that the
HyperLight-G be
powered from an
uninterruptible power supply
or battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30A/m
3 & 30A/m
Power frequency magnetic
fields should be that of a
typical commercial or
hospital environment.
Guidance and Manufacturer’s Declaration Electromagnetic Emissions
The HyperLight-G is intended for use in the electromagnetic environment specified below.
The customer or user of the HyperLight-G should ensure that it is used in such an
environment.
Immunity
test
IEC 60601 test
level
Compliance level
Electromagnetic environment -
guidance