11 EMC Tables
Page 38 / 42
Conducted RF
IEC
61000-4-6
Radiated RF
IEC
61000-4-3
3 Vrms
150 kHz to 80
MHz
6Vrms in ISM
bands between
0,15 MHz and
80 MHz
3 V/m
80 MHz to 2,7
GHz
(V1)=3Vrms
(E1)= 6Vrms in ISM
bands
(E1)=3V/m
Portable and mobile
communications equipment
should be separated from the
HyperLight-G by no less than the
distances calculated/listed below:
D=(3.5/V1)(Sqrt P)
150kHz to 80MHz
D=(3.5/E1)(Sqrt P)
80 to 800 MHz
D=(7/E1)(Sqrt P)
800 MHz to 2.5 GHz
where P is the max power in watts
and D is the recommended
separation distance in meters.
Field strengths from fixed
transmitters, as determined by an
electromagnetic site survey,
should be less than the
compliance levels (V1 and E1).
Interference may occur in the
vicinity of equipment containing a
transmitter.
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and D is
the recommended separation
distance in meters (m).