Attention: Store the sensors only in the original
container and within the specied temperature
range of -18°C to +8°C. Only remove the sensors
for the purpose of testing. Close the container
again immediately. Never leave open! Protect
against solar radiation, prolonged temperatures
above 25°C and moisture. Temperatures outside
the specied range during transportation and
storage may inuence the function of the sensors
and lead to faulty results.
Sensors, once removed from the container, must
be used immediately. The sensors must be kept at
room temperature with the measuring device at
least 20 minutes (at least two hours when stored
in the freezer) in the closed container prior to
testing. The expiration date of the sensors on the
label of the sensor container must be observed.
Following the initial opening of the container, the
sensors must be used within 90 days.