32
Technical specication
Parameter Specication
Type of device
Sample material
Sample volume
Measurement range
Precision
Measurement principle
Test duration
Measurement temperature range
Humidity
Storage temperature
Hematocrit range (Hct)
Max. elevation for use
Data storage
Data transfer
Power supply
Battery life
Device dimensions
Device weight
Lactate Scout+ handheld device for measuring blood lactate
(model from SN 0000101115)
Fresh capillary whole blood
0.2 μl
0.5 mmol/L – 25.0 mmol/L
± 3 % (minimal standard deviation: ± 0.2 mmol/L)
for Hct range 35 – 50%
± 4 % (minimal standard deviation: ± 0.3 mmol/L)
for extended Hct range
Enzymatic amperometric determination of lactate using
lactate oxidase
10 seconds
+5 °C - +45 °C
10 - 85% relative humidity (up to 2 minutes for sensors)
-18°C - +8°C (sensors in container) or -20°C - +50°C (device)
20 – 70 %
up to 4,000 metres
250 test values with date, time and step test allocations
Integrated Bluetooth® module (Class II module BT 1.2.)
2 x 1.5 V AAA / LR03 batteries
At least 1000 tests
91 mm (L) x 55 mm (B) x 22 mm (H)
80 g (with batteries)