31
Turn on Device fails to turn on Replace batteries. Press scroll wheel
> 2 seconds. If the problem persists
please contact LactateScout+ Service.
No test
readiness
No drop symbol displayed
Sensor not inserted properly
Insert sensor until it is noticeable in
place with the black contact surface
facing up.
Signal tones No acustic signal
Volume was muted in the volume setting menu
Select and activate the volume under
settings
Display Display shut off
Happens automatically after 5 minutes of inactivity
Poor display contrast
Too low ambient temperature or empty batteries
Display elements missing
Defective electronics or mechanical damage
Restart the device
Ensure ambient temperature in operating
range, replace batteries
Please contact Lactate Scout+ Service
Test values Values appear to high/too low Repeat the test using new sensor and
new coding, follow the instructions for
performing the test carefully. Perform
functionality test. In case of Er4 message
contact Lactate Scout+ Service.