Section 1. Getting Started
Clinical settings
Cartridge
The dose range depends on the type of cartridge selected containing a specified concentration of
Saizen
®
[somatropin (rDNA origin) for injection]. The following click.easy
®
cartridge concentrations
are available:
Cartridge type A (concentration of 8 mg/mL):
range can vary between 0.5 mg to 6.4 mg in increments of 0.01 mg.
Cartridge type B (concentration of 5.83 mg/mL):
range can vary between 0.15 mg to 4.66 mg in increments of 0.01 mg.
Cartridge type C (concentration of 1.5 mg/mL):
range can vary between 0.15 mg to 1.2 mg in increments of 0.01 mg.
IMPORTANT: The cartridge type should be selected according to your prescription.
To change the Cartridge setting:
1. Open the “Clinical settings” menu and
enter the PIN code (see page 14) (21a).
2. Use the key to highlight “Injection
settings”.
3. Select “Change to open the “Injection
settings” menu. “Cartridge” will
be highlighted.
4. Select “Change and the “Cartridge
menu will open (21b).
5. Use the key to highlight Cartridge “A”, “B”, or “C” (21c).
6. Select “OK”.
Clinical settings
Dose settings
Injection settings
Cartridge expiration
Dose log
Change Back
Injection settings
Change Back
Cartridge
Needle type
Cartridge
OK Cancel
A
B
C
21a
21b
21c
21