Section 4. Technical Information about easypod
Recommended separation distances between
portable and mobile RF communications equipment and easypod
easypod is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The user of easypod
can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and easypod as recommended below, according to the maximum output power of the communications
equipment.
Separation distance according to frequency of transmitter (m)
Rated maximum output of
transmitter
W
dP
=
3,5
V
1
[]
150 kHz to 80 MHz
dP
=
3,5
E
1
[]
80 kHz to 800 MHz
dP
=
7
E
1
[]
800 kHz to 2,5 GHz
0,01 Not applicable 0,04 0,07
0,1 Not applicable 0,11 0,22
1 Not applicable 0,35 0,7
10 Not applicable 1,11 2,21
100 Not applicable 3,5 7
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.