Table 2-2: Troubleshooting (continued)
Error Recommended action(s)
Configuration changed • One or more parameters have been modified in the meter's
configuration,
— Collect an Audit log using Daniel MeterLink in order to see what
configuration parameters changed and when they changed.
— Run the Tools → Edit/Compare Configuration utility and select
Write All or Write Checked values to write the changes to the
meter.
— Save the configuration file.
Configuration lost • The meter configuration has reset to default values and the meter is
not configured correctly to measure flow and the meter has
performed a Cold Start.
— Unless the Cold Start occurred after upgrading firmware,
replace the CPU board.
— If the cold start occurred after a firmware upgrade, fully re-
configure the meter from a previously saved configuration
using the Tools → Edit/Compare Configuration in Daniel
MeterLink.
Electronics Temperature
is Out of Nominal Range
• Temperature of the internal electronics is out of nominal operating
range (below -40 °C or above 100 °C) which could lead to a system
failure.
— Attempt to warm or cool the meter electronics housing.
— If the electronics is mounted to the meter and the process fluid
in the meter is over 60 °C, you must remote mount the
electronics off of the meter body.
— Collect a Maintenance log using Daniel MeterLink while the
meter is experiencing the issue and contact your Daniel service
representative.
Maintenance and Troubleshooting manual Troubleshooting
P/N 3-9000-791 August 2021
Maintenance and Troubleshooting manual 21