Note
Both the laser and sensor should be at the same height, as low as
possible, yet just high enough for the beam to clear the coupling flange.
They should also visually appear to be rotationally aligned to each other.
Horizontal alignmentFigure 2-3:
3. Make final adjustments, loosening the brackets slightly if necessary, then
rotating them and re-tightening.
2.4 Adjust the laser beam
1. Turn on the laser and sensor.
2. With the sensor lens covered, let the laser beam strike the center of the
sensor dust cap.
3. Slide the sensor dust cap to open the lens.
4. Observe the four laser beam adjustment LEDs on the sensor while
adjusting the laser beam using the vertical and horizontal beam
positioning thumbwheels on the laser.
Use the thumbwheels to adjust the horizontal and vertical laser beam
angles.
5. Continue with the adjustment until all four laser beam adjustment LEDs
on the sensor are blinking green once every second.