[76]
Figure 9-9: Measurement
Target material Material of the measuring object entered in the configuration (see
Section 8.5.4
).
Detuning
Displays the selected detuning level (see
Section 8.5.4
).
Extended range
factor
Measuring range
min.
Measuring range
max.
Frequency
Configured factor for the measuring range extension (see
Section 8.5.4
).
The beginning of the configured measuring range depends on the
selected sensor type and the factor for the measuring range
extension (see
Section 8.5.4
).
The end of the configured measuring range depends on the
selected sensor type and the factor for the measuring range
extension (see
Section 8.5.4
).
This graphic object displays the frequency of the oscillator circuit of the converter. The
frequency depends on the connected sensor type, the cable length of the sensor, and the
selected detune frequency (Measurement > Detune frequency).
Figure 9-10: Frequency
Sensor
This graphic object displays general sensor data.