Configuration
[64]
The measured voltage (approximately -10 V) is entered into the related field of
column [V].
9. With the micrometer screw, adjust the distance between measuring object and
sensor head to the minimum gap stated in row D of
Figure 8-27
.
10. Click the measuring button in row D to measure the output voltage of the
converter.
This process can take several minutes.
The measured voltage (approximately -2 V) is entered into the related field of
column [V].
11. Click Next.
12. Click Next to send the calibration table to the EZ 1000.
CAUTION!
Connections to external devices may be interrupted when sending configurations.
The output voltage of the EZ 1000 is set to > -1.5 V during this process.
The calibration and sending progress is displayed.
13. Click Next.
14. Click Finish to exit the wizard.
15. Click Close configuration to exit the configuration editor. Do not click Send & close.
8.6 Send and Reload a configuration
8.6.1 Send a configuration
CAUTION!
Connections to external devices may be interrupted when sending configurations.
The output voltage of the EZ 1000 is set to > -1.5 V during this process.
The configuration is generally sent during the calibration process. Use Send & close to send
minor changes to the configuration or configuration files loaded into Machine Studio to
the converter.
Procedure
1. Ensure that there is an online connection between the EZ 1000 converter and
Machine Studio.