[78]
• Factory def.: Converter has the factory default configuration
(delivery state).
• User def.: Converter has a user defined configuration.
Cal. type
Indicates the calibration type.
• Not available: Calibration state is not available.
• Auto: The converter has been calibrated with the Automatic
calibration function.
• Easy: The converter has been calibrated with the Offline
calibration function.
• Manual: The converter has been calibrated with the Manual
calibration function.
• Multipoint: The converter has been calibrated with the Multipoint
calibration function.
See
Section 8.5.5
and
Section 8.7
for details.
Gain Value which has been determined for the gain during the calibration
process.
Offset Value which has been determined for the offset during the
calibration process.
Sensor
frequency offset
Value which has been determined for the sensor frequency offset
during the calibration process.
9.3 Live data
Live data displays continues online data. The trend diagram shows the last five minutes. The
trend Voltage displays the DC voltage part of the converter output signal. The trend
Frequency displays the oscillator frequency of the converter.