Configuration
[63]
Figure 8-27: Multipoint calibration
A.
Buttons for measurement of converter output voltage.
B.
Calibration point for the end of the measuring range (greatest distance between sensor and
measuring object).
C.
Calibration point for the center of the measuring range.
D.
Calibration point for the beginning of the measuring range (smallest distance between sensor
and measuring object).
E.
Indication of the current output voltage of the connected converter.
F.
Calibration curve as a result of the calibration points.
Rows C and D are disabled until the measurement for the start point has been made
(row B).
6. Click the measuring button in row B to measure the output voltage of the
converter.
This process can take several minutes.
The measured voltage (approximately -18 V) is entered into the related field of
column [V].
7. With the micrometer screw, adjust the distance between measuring object and
sensor head to the center of the measuring range stated in row C of
Figure 8-27
.
8. Click the measuring button in row C to measure the output voltage of the
converter.
This process can take several minutes.