Technical data
[93]
First meter of EZ 108x at temperature of EZ 108x.
No ambient temperature change of EZ 1000 during 1 hour before measurement,
power and sensor connected to EZ 1000 for at least 10 minutes before
measurement and calibration.
WayCon reference target material is 42CrMo4 (AISI 4140), unless otherwise stated.
Note
For measurements that require a higher measuring accuracy, factory calibrated measuring chains are
available.
Calibration methods Offline (Easy) and Automatic typical values
The values of
Table 11-3
are typical values for measuring chains, calibrated with calibration
method Offline (Easy) or Automatic.
Accuracy of non-API 670 conform measuring chains
Add the measurement inaccuracies in
Table 11-4
to the values in
Table 11-3
to get the
accuracy of non-API 670 conform measuring chains.
Table 11-4: Additional inaccuracies DSL
Parameter
Additional
Measurement
Inaccuracy (DSL),
Test range
Additional
Measurement
Inaccuracy (DSL),
Operating range
Remarks
Other cable lengths
0.25%
0.75%
Other target materials
1.25%
1.25%
Material selected from
Machine Studio material
table
Extended measuring
range: 3 mm
(Extended range factor:
1.5)
Manual calibration
1.25%
1.5%
Automatic calibration
1.25%
1.5%
Multipoint calibration
0.25%
0.5%
Extended measuring
range: 4 mm
(Extended range factor:
2.0)
Manual calibration
3.75%
4.0%
Automatic calibration
3.75%
4.0%
Multipoint calibration
0.5%
1.0%