Maintenance (continued)Table 2-2:
Daniel MeterLink utility Diagnostics Action(s)
MeterLink Tools Menu Frequency output Run the Frequency Outputs
test
If the output reads zero,
you may require a pull up
resistor 1.2kOHM, 0.5W.
Analog outputs Run Analog Outputs test
Verify outputs are within
4mA-20mA range
- 0% = 4mA
- 25% = 8mA
- 50% = 12mA
- 75% = 16mA
- 100% = 20mA
Digital outputs Run the Digital Outputs
test.
Digital Output Content is in
relation to frequency validi-
ty and flow direction con-
figuration and polarity.
Meter Electronics Acquisition Module com-
munications error
Check firmware revision
and upgrade if necessary
using Daniel Meter-
Link Tools>Program
Download.
If LED 4 and LED5 are not
Solid Green check intercon-
nects cable between CPU
Acquisition Module. Check
IS Barrier Module or Acquis-
ition Module.
Troubleshooting
32 Gas Ultrasonic Flow Meters