Maintenance (continued)Table 2-2:
Daniel MeterLink utility Diagnostics Action(s)
MeterLink Logs/ Reports Menu Meter performed a Warm
Start or a Warm start re-
quired
Meter performed a Warm
Start:
- Collect an Archive event
log (Audit log) using
Daniel MeterLink to
view configuration pa-
rameter changes and
when they changed.
Warm start is required:
- When you make
changes to the trans-
ducer characteristics,
sample rates, the de-
vice number, or a Mod-
bus map file.
MeterLink Tools>Edit/
Compare Configuration
Menu
Meter performed a Cold
Start
The meter configuration
has reset to default values
and the meter is not config-
ured correctly to measure
flow.
Unless the cold start occur-
red after upgrading firm-
ware, you may need to re-
place the CPU Module.
If the Cold Start occurred
after a firmware upgrade,
you must reconfigure the
meter from a previously
saved configuration file us-
ing the Edit>Compare
Configuration screen.
Then clear the latched
alarm on the Status Sum-
mary page.
MeterLink Logs/ Reports Menu Power failure If this was a known power
fail or restart of the meter
just acknowledge this
alarm on the Status Sum-
mary page.
If this was an unexpected
restart of the meter, verify
the integrity of the power
to the meter and make sure
that the voltage level is in
the range of 11-36 VDC at
the meter.
Troubleshooting
Maintenance and Troubleshooting manual 33