24 Micro Motion
®
Model D and DT Sensors Instruction Manual
Wiring continued
Model D or DT sensor (except D600) wiring to RFT9712 transmitter
Model D sensor or DT (except D600) wiring to Model 1700 or 2700 transmitter
Model D sensor wiring (except the D600) to IFT9701* and Model 5300 transmitters
9 7531
86420
Brown
Red
Clip drain wire back
Green
White
Clip drain wire back
Blue
Gray
Clip drain wire back
Orange
Violet
Yellow
Clip drain wire back
Model D or DT
sensor terminals
Flowmeter
cable
RFT9712
terminals
Brown
Red
Green
White
Blue
Gray
Orange
Violet
Yellow
Black
(Drains from all
wire sets)
Brown
Orange
Green
Violet
White
Gray
Blue
Yellow
Red
Black (Drains)
Maximum cable length 1000 ft. (300 m)
Prepare cable in accordance with the instructions
that are shipped with the cable
Brown
Red
Orange
Yellow
Green
Blue
Violet
Gray
White
For DT sensor junction box
information, see page 20.
Brown
Red
Clip drain wire back
Green
White
Clip drain wire back
Blue
Gray
Clip drain wire back
Orange
Violet
Yellow
Clip drain wire back
Model D or DT
sensor junction box
terminals
Flowmeter
cable
Field-mount Model
2700 terminals
Brown
Red
Green
White
Blue
Gray
Orange
Violet
Yellow
Black
(Drains from all
wire sets)
Brown
Violet
Ye l l o w
Maximum cable length 60 ft. (20 m)
Prepare cable in accordance with the
instructions that are shipped with the cable.
Do not allow shields to contact sensor
junction box.
Blue
Gray
Orange
Red
Green
White
Ground screw
Black
Brown
Red
Orange
Ye l l o w
Green
Blue
Violet
Gray
White
10
11
12
GND
Brown
Red
Orange
Yellow
Green
Blue
Violet
Gray
White
Brown
Red
Clip drain wire back
Green
White
Clip drain wire back
Blue
Gray
Clip drain wire back
Orange
Violet
Ye l l o w
Clip drain wire back
Flowmeter
cable
Brown
Red
Green
White
Blue
Gray
Orange
Violet
Yellow
Black
(Drains from all
wire sets)
Maximum cable length 1000 ft. (300 m)
Prepare cable in accordance with the instructions
that are shipped with the cable
Black (Drains,
remote-mount only)
Brown
Red
Orange
Yellow
Green
Blue
Violet
Gray
White
Model D sensor
terminals
* Model D600 and DT sensors cannot be used with IFT9701 transmitters
IFT9701 or
Model 5300 terminals

Other manuals for Emerson Micro Motion

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Micro Motion and is the answer not in the manual?

Emerson Micro Motion Specifications

General IconGeneral
CategoryMeasuring Instruments
ManufacturerEmerson
BrandMicro Motion
TypeCoriolis Flow Meter
Output4-20 mA, HART, Foundation Fieldbus, Profibus, Modbus
Power Supply24 V DC
MaterialStainless Steel, Hastelloy
Process ConnectionFlanged, Threaded
CertificationsATEX, IECEx, CSA, SIL
Pressure RatingVaries by model; up to 10, 000 psi (690 bar)
Temperature RangeVaries by model; -400 to 750°F (-240 to 400°C)