4 Set the switches
The transmitter’s default configuration sets the alarm condition to
High (HI) and the security to Off.
Procedure
1. If the transmitter is installed, secure the bus and remove
power.
2. Required: Remove the transmitter cover opposite the field
terminal side. Donot remove the instrument covers in
explosive environments when the circuit is live.
3. Slide the Security and Alarm switches into the preferred
position by using a small screwdriver.
Note
The Security switch will need to be in the off position in order
to make any configuration changes.
4. Required: In order to meet explosion-proof requirements,
reinstall the housing cover and tighten so the cover is fully
seated with metal to metal contact between the housing and
cover. After the cover is seated properly, replace the flathead
screw located on the bottom of the housing cover.
January2024 Quick Start Guide
Quick Start Guide 13