Emerson Process Management GmbH & Co. OHG S-9
X-STREAM XE
Short Form Manual
HASXEE-SFM-HS
05/2017
S
Safety Instructions
Safety Instructions
HEAVY INSTRUMENT
The models intended for outside and wall mounted use (X-STREAM XEF and
XDF) weigh between 25 kg (57 lb) and 45 kg (99 lb), depending on version
and options installed.
Two people and/or lifting equipment is required to lift and carry these units.
Take care to use anchors and bolts specied to be used for the weight of
the units!
Take care the wall or stand the unit is intended to be installed at is solid
and stable to support the weight!
HIGH TEMPERATURES
Hot parts may be exposed when working on photometers and/or
heated components in the unit.
CRUSHING HAZARD
Take care of crushing hazard when closing the front door of analyzer eld
housings!
Keep out of the closing area between enclosure cover and base!
OPERATION AT LOW TEMPERATURES
When operating an instrument at temperatures below 0 °C (32 °F), do
NOT apply gas nor operate the internal pump before the warmup time has
elapsed!
Violation may result in condensation inside the gas paths or damaged
pump diaphragm!

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Rosemount Analytical X-STREAM Enhanced Series and is the answer not in the manual?

Emerson Rosemount Analytical X-STREAM Enhanced Series Specifications

General IconGeneral
BrandEmerson
ModelRosemount Analytical X-STREAM Enhanced Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals