Emerson Process Management GmbH & Co. OHG 2-3
X-STREAM XE
Short Form Manual
HASXEE-SFM-HS
05/2017
2
Installation
2.3 Technical Data
2.3.1 Common Technical Data
Before intending to start to install the ana-
lyzer, verify the site of installation meets the
specications, given in the following sections!
2.3 Installation - Technical Data
Site of installation
Humidity
(non-condensing)
< 90 % RH at +20 °C (68 °F)
< 70 % RH at +40 °C (104 °F)
Degree of pollution 2
Installation category II
Elevation 0 to 2000 m (6560 ft) above sea level
Ambient atmosphere
Units may not be operated in corrosive,
ammable or explosive environments
without additional safety measures.
Certication
Electrical safety
CAN / USA
CSA-C/US, based on CAN/CSA-C22.2 No.
61010-1-04 / UL 61010-1, 2nd edition
Europe
CE, based on EN 61010-1
Electromagnetic
compatibility
Europe CE, based on EN 61326
Australia C-Tick
others NAMUR