EasyManuals Logo

Emerson Rosemount MultiVariable 3051S Series User Manual

Emerson Rosemount MultiVariable 3051S Series
52 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #21 background imageLoading...
Page #21 background image
Figure 6-4: Fluid Selection Tab
Note
The following example will show a flow configuration for the database gas air
used with a Rosemount 405C Conditioning Orifice Plate as the primary
element. The procedure to set up any other fluid with any other primary
element will be similar to this example. Natural gases, custom liquids, and
custom gases require additional steps during the configuration. See the
Rosemount 3051SMV Reference Manual for more information.
Procedure
1. Engineering Assistant may open to the Preferences tab. Using the tabs
at the top of the screen, navigate to the Fluid Selection tab.
2. Expand the Gas category (click on the + icon).
3. Expand the Database Gas category.
4. Select Air from the list of database fluids.
5. Enter the Nominal Operating Pressure, select the Enter or Tab key.
6. Enter the Nominal Operating Temperature, select the Enter or Tab key.
Engineering Assistant will automatically fill in suggested operating
ranges, as shown in Figure 6-4. These values may be edited as needed
by the user.
7. Verify the Reference/Atmospheric Conditions are correct for the
application. These values may be edited as needed.
Note
Reference pressure and temperature values are used by Engineering
Assistant to convert the flow rate from mass units to mass units
expressed as standard or normal volumetric units.
October 2018 Quick Start Guide
Quick Start Guide 21

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Rosemount MultiVariable 3051S Series and is the answer not in the manual?

Emerson Rosemount MultiVariable 3051S Series Specifications

General IconGeneral
TypeMultiVariable Transmitter
Accuracy±0.04% of span
Ambient Temperature Limit-40 to 185 °F (-40 to 85 °C)
ApprovalsFM, CSA, ATEX, IECEx
ServiceLiquid, Gas, Steam
Warranty5 years
Process Temperature Range-40 to 257°F (-40 to 125°C)
Response Time100 ms
Power Supply10.5 to 42 V DC
Communication ProtocolsHART, FOUNDATION Fieldbus, PROFIBUS
Housing MaterialAluminum
DisplayOptional LCD
Process Connection1/4" NPT
Electrical Connection1/2–14 NPT, M20 x 1.5
Enclosure RatingNEMA 4X, IP66
Measurement TypesDifferential Pressure, Static Pressure, Temperature

Related product manuals