eMKW-TD0054-21- Page 72
This report page shows the test frequencies as well as the Amplitude Error.
The Amplitude Error is the discrepancy or difference between the calibrated level of the KUDUwave’s
air conduction output and the actual air conduction output, which has just been measured by the X-
check. Ideally all of these values should be zero.
If you would like a PDF report, scroll down and click the button. This will generate
and open a PDF, like the one shown here. Depending on the version of X-Check you are running. The
PDF report is automatically saved in the C:\Program Files (x86)\eMoyoDotNet\Pdf folder on your
computer, or C:\Program Files (x86)\eMoyoDotNet\eMOYO\Data\Apps\KUDUwave\X-check_tests.