eMKW-TD0054-21- Page 73
X-check: Abnormal Test
If the test fails, a message will list common reasons which cause X-check to fail alternatively to your
KUDUwave being out of calibration.
As an example, bumping or moving the KUDUwave while the X-check is running will prevent it from
accurately measuring the air conduction output and cause the test to fail. Follow the instructions and
click the “Retest” button for the failed side.
You can also click the “Detail view” button to go to the Detail page and get more information about why
the test failed. Failed frequencies are shown in red along with a reason regarding why the test failed.