You will need a lancing device in order to collect a blood
sample.
You may use the lancing device contained in the Bravo™ Blood
Glucose Monitoring System or any other medically approved
lancing device.
patient. It must not be used on more than one person
due to risk of infection.
necessary, a small amount of alcohol on a soft cloth or
tissue may be used.
USING THE LANCING DEVICE
Repeated puncturing at the same sample site may
cause pain or skin calluses (thick hard skin). Choose
a different site each time you test.
Note
To avoid infection when drawing a sample, use
a lancet only one time, and:
Caution
!
Adjustable Tip
Lancet Holder
Sliding Barrel
Release Button
Ejection System
Lancet
Comparing the Control Solution Test Results
The test result of each control solution should be within the
range printed on the label of the test strip vial. Repeat the
control solution test if the test result falls outside of this range.
Out of range results may occur due to the following factors:
If results continue to fall outside the range printed on the
test strip vial, the Bravo™ Test Strip and Meter may not be
working properly. Do not use your system and contact
Endomedical™ sales representative.
not shaken well,
control solution were exposed to high
or low temperatures,
solution was not discarded or the tip
of the bottle was not wiped clean,
properly.
Repeat the control
solution test by
referring to the
“Notes” on page
17.
expiration date printed on the bottle,
discard date (the date the bottle was
opened plus three (3) months),
contaminated.
Discard the used
control solution
and repeat the test
using a new bottle
of control solution.
Situations Actions