Endomedical Bravo - Understanding Error Messages; Error Code Explanations and Actions

Default Icon
23 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
BLOOD GLUCOSE MONITORING SYSTEM
40
www.endomedical.ca
41
A used test strip
was inserted.
Repeat the test with
a new test strip.
The blood or control
solution sample was
applied before the
symbol
appeared.
Repeat the test with
a new test strip and
wait until the
symbol appears
before applying the
blood or control
solution sample.
The temperature
during the test was
above or below
the operating
range.
Move to an area
where
the
temperature is
within the
operating range (10
~ 40°C / 50 ~ 104°F)
and repeat the test
after the meter
and test strips
have reached a
temperature within
the operating
range.
UNDERSTANDING ERROR AND OTHER MESSAGES
MESSAGE WHAT IT MEANS WHAT TO DO MESSAGE WHAT IT MEANS WHAT TO DO
WHAT TO DO
The blood sample
has abnormally
high viscosity or
insufficient volume.
Repeat the test
after inserting
a new test strip.
This error message
may appear when
the wrong blood
glucose test strip
is used instead
of Bravo™ blood
glucose test strip.
Repeat test after
inserting a Bravo
test strip.
There is a problem
with the meter.
Do not use the
meter.
Contact
your authorized
Endomedical™ sales
representative.
If the error messages persist, contact your
authorized Endomedical™ sales representative.
Note