BLOOD GLUCOSE MONITORING SYSTEM
44
www.endomedical.ca
45
Precision: The precision studies were performed in a laboratory
using Bravo™ BGM Systems.
Within Run Precision
*Blood avg. 2.8 mmol/L SD = 0.1 mmo/L
*Blood avg. 5.1 mmol/L SD = 0.2 mmo/L
*Blood avg. 7.8 mmol/L CV = 2.4%
*Blood avg. 11.3 mmol/L CV = 2.9%
*Blood avg. 19.2 mmol/L CV = 2.9%
Total Precision
*Control avg. 2.6 mmol/L SD = 0.1 mmol/L
*Control avg. 7.6 mmol/L CV = 3.1%
*Control avg. 18.7 mmol/L CV = 3.5%
This study shows that there could be variation of up to 3.5%.
PRECISION