EasyManua.ls Logo

Endress+Hauser cerabar S PMC Series - Mounting Instructions with Diaphragm Seal (PMC 631, PMP 635)

Endress+Hauser cerabar S PMC Series
68 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.2 Mounting instructions with diaphragm seal (PMC 631, PMP 635)
Cerabar S
with diaphragm seal
– PMC 635
– PMP 635
The Cerabar S with diaphragm seal is
screwed in, flanged or clamped,
depending on the type of diaphragm seal.
The protective cap of the diaphragm
seal should only be removed just before
mounting in order to protect the
diaphragm.
The membrane of the diaphragm seal of
the Cerabar S must not be pressed in or
cleaned with pointed or hard objects.
The diaphragm seal and the pressure
sensor together form a closed and
calibrated system which is filled with
filling fluid through a hole in the upper
part. The following rules should be
observed:
This hole is sealed and not to be
opened.
The device may only be rotated on
the surface provided on the
diaphragm seal instrument and not on
the housing.
Level measurementFor level measurement the Cerabar S must
always be installed below the lowest
measuring point.
The instrument should not be installed
within the flow of material, in the outlet
of the tank or at a place in the tank
where pressure pulses from agitator
blades may occur.
Calibration and function testing are
easier to carry out if the Cerabar S is
mounted downstream from a shut-off
valve.
Mounting with
temperature spacers
The use of temperature spacers is
recommended for continuous extreme
product temperatures.
Note when mounting that the
temperature spacer increases the
maximum height by 100 mm.
Due to the hydrostatic column in the
temperature spacer, the increased
height also causes a zero point shift of
approx. 10 mbar.
See Chapter 5.1, pages 26 and 27 for
information concerning position
calibration (display only) or zero-point
correction.
0 - 10 bar
BA187Y16
Figure 2.6
When screwing in the Cerabar S
with diaphragm seals, turn by the
diaphragm seal only, not by the
housing.
0 - 10 bar
BA187Y12
0 - 10 bar
A + 100
100
BA187Y13
Figure 2.7
Information on the maximum
installation height A is given on
pages 57…58
Cerabar S 2 Installation
Endress+Hauser 13

Related product manuals