3.4 Operation with Commuwin II
When operating the Commuwin II display and operating program, the Cerabar S is set
and and operated either using:
Matrix mode or
Graphic mode.
The appropriate server (e.g. HART or ZA 672) must therefore be activated. A description
of the operating program Commuwin II is to be found in Operating Instructions BA 124F.
Note!
You can obtain the current device description (DD) either from your local
Endress+Hauser Sales Center or via the Internet (http://www.endress.com Products
Product Portfolio Process Solutions Commuwin II Updates/Downloads).
Matrix mode
(Menu Device)
You can access the extended functions of the Cerabar S, such as level measurement,
using the "Device/Parameter Matrix" menu.
Each row is allocated to a particular function.
Each field sets or displays one parameter.
Enter the setting parameters in the appropriate fields and confirm by pressing .
Graphic mode
(Menu Device)
Commuwin II offers graphic examples of certain configuration procedures which you can
access from the "Device/Graphics" menu. There you can directly modify parameters and
confirm by pressing ↵.
BA187E33
Figure 3.2
Menu "Device/Parameter matrix"
in Commuwin II
BA187E32
Figure 3.3
Menu "Device/Graphics"
in Commuwin II
Note!
3 Operation Cerabar S
20 Endress+Hauser