16 Endress+Hauser
Reject entry.
If this symbol is selected, the entry is rejected and you quit editing mode. The previously set text
remains.
Jump one position to the left.
If this symbol is selected, the cursor jumps one position to the left.
Delete backwards.
If this symbol is selected, the character to the left of the cursor position is deleted.
Delete all.
if this symbol is selected, the entire entry is deleted.
5.4 Operating functions
The operating functions of the transmitter are organized into the following menus:
Display Settings for the device display: contrast, brightness, time for alternating measured values on the
display
Setup Device settings
A description of the individual settings is provided in the "Commissioning" section →  17.
Calibration Execution of the sensor calibration
A description of the functions for calibration is provided in the "Calibration" section.
Diagnostics Device information, diagnostics logbook, sensor information, simulation
5.5 Hold function
The hold function causes the current outputs and relay states to "freeze". This function can be
switched on and off manually (menu SetupManual hold). In addition, the hold function is
automatically activated during sensor calibration.
When the hold condition no longer applies, the hold function continues to be active for the
configurable hold release time. The hold release time is configured in the menu Setup
Extended setupSystemHold release.
The hold function does not affect the display of the measured value. The hold symbol is also
displayed after the measured value.