Commissioning CM14
18 Endress+Hauser
5. Press E to open the Extended Setup menu; System is displayed.
6. Press E to open the System menu.
7. Press + repeatedly until Access code is displayed.
8. Press E to open the setting for access protection.
9. Set the code: press the + and - buttons to set the desired code. The access code is a four-
digit number. The corresponding position of the number is displayed in plain text. Press
E to confirm the value entered and go to the next position.
10. Confirm the last position of the code to exit the menu. The full code is displayed. Press +
to scroll back to the last item of the x Back submenu and confirm this item. By
confirming the point, the value is adopted and the display returns to the Setup level.
Again select the last parameter x Back to also exit this submenu and return to the
measured value/channel display level.
Once access protection has been successfully activated, the lock symbol appears on the
display.
Access to the calibration function can also be blocked by a code. The same procedure is
required here as for locking the setup. To activate it, however, press + repeatedly in step
7 until Calib Code is displayed.
The x Back item at the end of every picklist/menu item takes the user from the submenu
to the next menu level up.
If access protection is enabled, the device locks automatically after 600 seconds without
operation. The display switches back to the operating display.
To enable the setup, set the setup access code in the System Setup to 0000 or delete the
code by pressing C.
If you lose/misplace the code, a reset can only be performed by the Service Department.
6.4 Configuration of the device (Setup menu)
You can access the main menu by pressing the 'E' key during operation. Navigate through the
available menus with the '+' and '–' keys. When the desired menu is displayed, press the 'E' key
to open the menu. Use the "x Back" option, which can be found at the bottom of each menu/
submenu, to move up one level in the menu structure.
The Setup menu contains the most important settings for the operation of the device.
Parameters Possible settings Description
Tag Customized text
Max. 16 characters
Use this function to enter the device tag.
Current range 4-20 mA
0-20 mA
Configuration of the measuring range for the
current output.

Other manuals for Endress+Hauser CM14