CM14 Troubleshooting
Endress+Hauser 33
Diagnostic
code
Message text Description
F100 Sensor comm.
No sensor communication.
Possible reasons:
No sensor connection
Faulty sensor connection
Short-circuit in the sensor cable
Short-circuit in the neighboring channel
Sensor firmware update canceled with an error
Remedy:
Check sensor cable connection
Check sensor cable for short-circuit
Replace sensor
Restart firmware update
Contact service
F130 Sensor supply
Sensor check.
Poor energy supply to sensor.
Remedy:
Check cable connections
Replace sensor
F142 Sensor signal
Sensor check.
No conductivity display.
Possible reasons:
Sensor in air
Sensor defective
Remedy:
Check sensor installation
Replace sensor
F143 Self test
Sensor self-test error.
Remedy:
Replace sensor
Contact service
F152 No airset
Sensor data.
No calibration data present
Remedy:
Carry out airset calibration
F523 Cell const.
Sensor calibration warning.
Invalid cell constant, max. range reached.
Remedy:
Recalibrate
Enter cell constant according to factory specifications
Replace sensor
F524 Cell const.
Sensor calibration alarm.
Min. possible cell constant undershot.
Remedy:
Recalibrate
Enter cell constant according to factory specifications